DC 16DEX
The DC 16DEX high vacuum pump has been realized in compliance with Directive 2014/34/EU (ATEX) and is designed specifically for ATEX Zone 2. It can work with clean air, flammable gas group IIA, IIB and IIC or small amounts of water vapor. The temperature of fluid, gas or steam processed by the pump and the environment in which it is installed must be between +12 ° C and +40 ° C.
The DC 16DEX vacuum pump is suitable for the evacuation of closed containers with a capacity allowing the minimum pressure to be reached in less than 20 minutes. As it has marking and certification for internal ATEX, it cannot be installed in areas classified as ATEX.
The DC 16DEX pump is equipped with:
- On/Off switch (1~ only)
- Cable without plug (1~ only)
- Thermal protector (130°C, 1~ only)
- Gas ballast
- Lifting and transport eyebolt
- Shock mounts
- Pack of SW100 oil
- Hydraulic non-return system
Main application fields: refrigerant charging machines, laser, leak detectors, freeze dryers, mass spectrometers and chromatography, laboratory autoclaves, microscopy, space simulators, under vacuum distillers, gel dryers, centrifuges and laboratory autoclaves, metallurgy, metallization, semiconductors.
Nominal capacity | Total final pressure | Stages | Motor power | |
m³/h 50 Hz / 60 Hz | mbar (Ass.) | kW 50 Hz / 60 Hz | ||
DC 16DEX | 17 / 21 | 0,005 | 2 | 0,55 / 0,66 |